ScanProof

A guardrail for chest X-ray classifiers. It runs four checks on every prediction and returns PASS, REVIEW or BLOCK along with the measurement that drove the verdict.

The problem

A two-class softmax splits its probability between the two classes it knows. Feed it something it has never seen and it does not get quieter, it just picks a side. Nothing in the output can mean "I don't recognise this".

My model, fine-tuned on pediatric chest films from one hospital, reports 86.0% mean confidence on adult films from another. On breast ultrasound it reports 95.8%, higher than on the X-rays it was actually trained for.

So ScanProof does not try to build a better confidence score. It runs four checks that fail for different reasons:

Check Asks Fails when
Typicality Has the model seen inputs like this? the embedding sits far from the training manifold
Stability Does the answer survive harmless edits? the label flips under a label-preserving perturbation
Agreement Do independently trained models concur? three checkpoints disagree
Confidence How far from the decision boundary? the margin is thin

There is a weighted 0-100 score for ranking, but what you act on is which check failed.

Results

Every figure below comes from make audit and make shift and is committed under artifacts/. pytest tests/test_claims.py parses them back out of this file and fails if they drift.

The deployment test

Train on children, run on adults. This is the ordinary way a deployed imaging model breaks: not a weird input, just somebody else's patients. Fine-tuned on pediatric films (ages 1-5, Guangzhou Women and Children's Medical Center), then run on adult frontal films from the NIH Clinical Center.

Measure Pediatric (n=624) Adult (n=484)
Model confidence 93.6% 86.0%
Embedding percentile 50.1 98.4
PASS rate 62.0% 4.8%

Confidence drops 7.6 points, nowhere near enough to notice in production. Typicality goes from the middle of the training distribution to its 98th percentile, and the guardrail withholds 95% of these inputs.

All three numbers are label-free. They are properties of the input and the model, so you can compute them on a live stream before anyone knows the right answer.

Sub-scores, pediatric to adult: typicality 0.878 to 0.116, agreement 0.801 to 0.619, confidence 0.892 to 0.759, stability 0.853 to 0.747. The embedding check carried this almost single-handed. The other three stay because they catch cases this arm does not contain, such as the Confident but fragile case in the demo deck, where typicality is fine and the label still flips.

The confound is controlled

The adult set is only available at 128 px unless you download 3.7 GB, so it gets resampled to 224. A control arm puts the pediatric films through that identical path. PASS rate moves 0.96 points, accuracy 1.1 points. Resolution is not what the study is measuring.

The negative result

On in-distribution data, plain confidence is a better error ranker than my composite score: AURC 0.0126 against 0.0175. Lower AURC is better, higher AUROC is better.

Signal In-distribution AURC Shift AUROC Worst case
Model confidence 0.0126 0.7479 0.203
Perturbation instability 0.0161 0.6941 0.000
ScanProof composite 0.0175 0.7955 0.383
Ensemble disagreement 0.0188 0.7113 0.065
Embedding percentile 0.0385 0.9592 0.000

I fixed the margins for "acceptable in both regimes" before running the study: within 0.01 AURC and 0.05 AUROC of the best signal in each. As measured, no signal clears both, mine included. I did not loosen them afterwards, and the failure is printed on the audit page rather than buried here.

What the evidence does support is narrower. Rescale each regime so the best signal scores 1 and the worst 0, then take the lower of a signal's two scores. That is the question a deployed system faces, since it commits to one number without knowing which failure shows up next. On that criterion the composite leads at 0.383 against 0.203, and nothing beats it on both axes.

That is worst-case coverage, not dominance. ScanProof does not beat confidence in-distribution. It covers a failure mode confidence cannot see.

In-distribution behaviour

On the 624-image held-out pediatric test split: accuracy 94.2%, AUROC 0.992, ECE 0.045 falling to 0.041 after temperature scaling. The PASS band reaches 97.9% accuracy at 61.1% coverage, and 28 of the 36 test errors land outside it.

Quick start

make setup       # .venv (Python 3.12) + python deps + npm install
make data        # PneumoniaMNIST + OOD probes (~250 MB, once)
make shift-data  # adult ChestMNIST arm (~1.4 GB, once)
make train       # fine-tune 3 members, calibrate, fit OOD stats (~15 min on Apple Silicon)
make audit       # select thresholds on val, evaluate on test
make shift       # the domain-shift study (~3 min)
make demo        # build the demo case deck
make build       # frontend
make serve       # http://127.0.0.1:8000

make reproduce runs all of it in order. Expect about 20 minutes from a clean checkout.

Before demoing, run make preflight. It re-analyses every shipped case live, diffs against the committed cache, and refuses placeholder artifacts. DEMO.md is the five-minute walkthrough.

For frontend work, make dev runs the API on :8000 and Vite on :5173 with a proxy. You need Python 3.10-3.13, Node 18+, and uv. MPS and CUDA are used when present; CPU works and is slower.

The demo path makes no network calls. Fonts are vendored, the frontend is served by the same FastAPI process as the API, and every model and dataset asset is local once make data and make train have run.

How the score is built

Four sub-scores in [0,1], combined with fixed weights from config.WEIGHTS:

Sub-score Weight Measures Zero when
stability 0.40 survival of the perturbation battery 25% of variants flip, or mean absolute delta-p at least 0.25
agreement 0.25 spread across three checkpoints sigma of P(pneumonia) at least 0.25; capped at 0.35 on a split vote
confidence 0.20 calibrated margin of the predicted class confidence at or below 0.50; saturates at 0.95
typicality 0.15 embedding distance from the training manifold at the 100th percentile; decay starts at the 90th

The weighted score is banded, then three gates can override it:

  • OOD gate. Embedding percentile at or above 99.5 forces BLOCK. The model has no comparable training examples, so its output is not interpretable at any confidence.
  • Fragility gate. A class flip at the mildest severity caps the verdict at REVIEW.
  • Split-vote backstop. A checkpoint disagreement can never reach PASS.

The gates override the weighted score deliberately. A failure this specific should not get averaged away by good numbers elsewhere.

The perturbation battery

Seven families at three severities each, 21 variants, all deterministic and all label-preserving.

Family Severities Stands in for
Brightness +12% / +24% / +36% exposure shift
Contrast -15% / -28% / -40% windowing
Gamma 1.25 / 1.55 / 1.90 display transfer curve
Detector noise sigma 0.02 / 0.045 / 0.08 sensor noise
Blur sigma 0.8 / 1.6 / 2.6 px focus or motion softness
Rotation 3 / 6 / 10 degrees patient or detector rotation
Resolution loss 0.70x / 0.50x / 0.35x lossy storage, lower-resolution acquisition

Rotation is reflect-padded, because black corners would be an out-of-distribution cue in their own right and would confound stability with typicality.

Thresholds

python -m scanproof.evaluate selects them on the validation split and writes artifacts/reliability_config.json. The test split is never used to choose anything.

  • review_threshold: the score quantile putting 12% of validation in BLOCK.
  • pass_threshold: the stricter of the lowest score whose PASS band reaches 99% selective accuracy, and the quantile capping PASS coverage at 70%.

The coverage cap matters. The ensemble nearly saturates validation, so an accuracy target alone is satisfied by passing everything, at which point the verdict stops discriminating. Selected values were PASS at or above 97.09 and REVIEW at or above 84.04, frozen before any test or shift arm ran.

The shift study

make shift runs four arms, ordered by distance from the training distribution:

Arm Data n Accuracy Confidence PASS rate
in_distribution PneumoniaMNIST test, native 224 624 94.2% 94.5% 61.1%
resolution_control the same films, resampled 624 93.1% 93.6% 62.0%
domain_shift ChestMNIST, adult, balanced 484 62.6% 86.0% 4.8%
wrong_modality BreastMNIST 156 n/a 95.8% 0.6%

Accuracy falls 30.5 points. Confidence falls 7.6. On breast ultrasound confidence does not just fail to drop, it rises to 95.8%, higher than on the adult films that were at least the right modality.

Nothing in the study re-tunes a threshold. scanproof.shift loads reliability_config.json and changes nothing in it.

Two caveats belong next to this. The composite is a worse shift detector than the raw embedding percentile, 0.7955 against 0.9592, because averaging four signals dilutes the one carrying the shift. And ChestX-ray14 labels are NLP-mined from free-text reports, so the 62.6% accuracy figure is indicative rather than a clean benchmark. Neither touches the headline, which is about the model's own confidence and needs no labels.

Layout

scanproof/      Python package: inference, reliability, evaluation, API
frontend/       Vite + React + TypeScript + Tailwind v4
tests/          pytest suite
artifacts/      generated JSON, committed
demo_cases/     PNGs + manifest with cached results, committed

reliability.py imports no torch and touches no disk, so the gate and arithmetic tests run in milliseconds against plain arrays. config.py holds the ensemble composition, weights and thresholds, which is the one place to audit the formula.

Run make test for the suite: perturbation determinism, severity monotonicity, structure preservation, sub-score arithmetic, every gate, and API behaviour in live and degraded mode.

Data and model sources

Asset Source License
PneumoniaMNIST, 5,856 pediatric chest X-rays medmnist.com, Zenodo 10519652 CC BY 4.0
ChestMNIST, NIH ChestX-ray14 adult films same CC BY 4.0
BreastMNIST, 780 breast ultrasound images same CC BY 4.0
ResNet-18 and DenseNet-121 ImageNet weights torchvision IMAGENET1K_V1 BSD-3-Clause

Datasets are fetched through the official medmnist package, which verifies MD5s against the Zenodo record. Nothing is scraped. MedMNIST images are de-identified and published by the dataset authors. No private or identifiable medical data is used anywhere in this repository.

PneumoniaMNIST derives from Kermany et al., Cell 176(2), 2018. ChestMNIST derives from Wang et al., CVPR 2017. Both via Yang et al., "MedMNIST v2", Scientific Data 10, 41 (2023). Temperature scaling follows Guo et al. (ICML 2017) and Mahalanobis OOD detection follows Lee et al. (NeurIPS 2018), both re-implemented from the published descriptions.

One thing worth knowing about the splits: MedMNIST builds PneumoniaMNIST's train and val splits from the source training collection and uses the source validation collection as test. So test is genuinely shifted, and the ensemble scores about 98% on validation but materially lower on test. That is what makes it a useful reliability benchmark, since there are real errors left to catch.

Limitations

  1. PASS is not a correctness guarantee. The PASS band is 97.9% accurate, so 8 of 381 passed cases are still wrong. The demo deck ships one on purpose.
  2. The data is 28x28-derived. PneumoniaMNIST is centre-cropped and resampled from full-resolution films, so fine detail and the periphery are gone before the model sees anything. None of this transfers to full-resolution DICOM without re-validation.
  3. Pediatric, single-source, binary. Real reporting is multi-label and multi-institution.
  4. Naturalistic corruption only. The battery says nothing about adversarial robustness.
  5. Single-layer Mahalanobis, weaker than multi-layer or ensembled OOD detectors. It needs no OOD data at fit time and runs in one forward pass, which is why I chose it.
  6. The weights are a judgement call. The 0.20/0.40/0.25/0.15 split was fixed a priori, and the shift study shows what it costs. A learned or regime-aware weighting would likely beat it, and it is the clearest next thing to build.
  7. The adult arm's labels are noisy, NLP-mined from free-text reports. No claim that matters rests on them.
  8. One shift axis, one direction. Pediatric to adult, one hospital to one other. Says nothing about scanner vendor, view position, or the reverse direction.
  9. Three checkpoints is a small ensemble, so disagreement is a noisy estimate from three votes.
  10. Thresholds come from 524 validation images and the adult arm is 484, so both carry real sampling variance. Bootstrap CIs are in shift_study.json.
  11. No temporal or site-level validation, no reader comparison, no prospective evaluation.